Remote Operation: What Front Panel Settings are Required to use the VXIPnP drivers for SourceMeter?

질문:

Remote Operation: What Front Panel Settings are Required to use the VXIPnP drivers for SourceMeter?

답변:

What Front Panel Settings are Required to use the VXI PnP drivers for the SourceMeter Instruments?


Selection and configuration of GPIB or RS-232 communication is important. Both are supported by the VXI PnP driver, however, the instrument must be configured from the front panel first.


Use the front panel buttons and select:

  • Menu
  • Communication
  • GPIB or RS-232

Instrument Settings

For GPIB:
For RS-232:

select the address (default is 24)

select SCPI protocol (not 488.1)

BAUD: 9600

BITS: 8

Parity: NONE

Terminator: LF

Flow-CTRL: XON-XOFF

Typical Resource Names

instrument at address 24

instrument on COM1

GPIB0::24::INSTR

ASRL1::INSTR

 

 

 

 

 

 

 


이 FAQ는 다음에 적용 됩니다:

제품 시리즈: 2400 시리즈 SMU

제품:

FAQ ID 69611