다운로드
다운로드

매뉴얼, 데이터 시트, 소프트웨어 등을 다운로드할 수 있습니다.

다운로드 유형
모델 또는 키워드

After Erasing the NVRAM+ I get an error E2211. What does the error indicate?

질문:

After Erasing the NVRAM+ I get an error E2211. What does the error indicate?

답변:

This is normal. To erase the NVRAM, the WAVEFORM and TRIGGER are held down until the power-up self-test is completed. The diagnostic system sees that two buttons are engaged. The diagnostic system reports this "error".

To continue, press Diagnostic Exit box on the display twice. Once the diagnostic system is completed, the graticule will be displayed. At the top, the following message with be displayed, "Teksecure Erase Memory Status: Erased, Instrument ID, on-time, and number of power-ups retained".


이 FAQ는 다음에 적용 됩니다:

제품 시리즈 없음

제품:

FAQ ID 57106

모든 FAQ 보기 »